Overview
The PM5020 Dual-Channel Optical Power and Energy Meter Console interfaces with one or two connected sensors to measure continuous wavelength and modulated light sources. This console is compatible with over 35 power, position, and energy sensors. When used with these sensors, the console will automatically detect and read the sensor calibration information stored in the sensor connector and combine this data with a user-set measurement wavelength to provide calibrated power or energy measurements. The console can also be used with third-party or custom sensors that meet the compatibility requirements outlined. Please note that sensors are not included with the PM5020 console.
Specifications
| Display | |
|---|---|
| Type | Graphical IPS LCD with Touch, 854 x 480 Pixels |
| Format | Numerical, Bargraph, Needle, Graph Power/Position/Temperature, Statistics |
| Size | 5.0" (4.32" x 2.43", 109.7 mm x 61.6 mm) |
| Update Rate | 25 Hz |
| Sensor Interface | |
| Compatible Sensors | Thorlabs' C-Series Photodiode, Thermal Power, Thermal Position & Power, and Pyroelectric Sensors Photodiodes up to 10 mA Thermopiles up to 1 V Pyroelectric Sensors up to 100 V |
| Channels | 2 |
| Input Connectors | Two DE-9 Female, Back Panel (Compatible with C-Series Connectors) |
| External Trigger | |
| Connector | BNC, Front Panel (TTL 5 V Max) 10-Pin PCB, Back Panel (TTL 5 V Max) |
| Analog Outputs | |
| Connector | Two BNC, Back Panel |
| Signal | CH1, CH2 Amplified, Corrected Input Signal |
| Voltage Range | -5 to 10 V; Programmable Gain and Offset |
| Bandwidth | Up to 1 kHz, Dependent on Sensor and Settings |
| Shutter Control | |
| Tyoe | Support for SH05R(/M) or SH1(/M) Optical Shutter with Interlock Input |
| Connector | 2 Channels: 2.5 mm Stereo Audio and 10-Pin PCB (Back Panel) |
| Digital I/O | |
| Type | 4 Configurable I/O Channels, 1 Trigger In/Output, I2C (LVTTL / TTL) |
| Connector | 10-Pin PCB, Back Panel |
| Fan Control | |
| Type | 2 Channels, Voltage or rpm Control (12 V, 0.25 A) |
| Connector | 10-Pin PCB, Back Panel |
| Power Rating | 12 V, 0.25 A |
| Interface | |
| Type 1 | USB2.0 |
| Connector | USB B, Back Panel |
| Type 2 | Ethernet |
| Connector | RJ45, Back Panel |
| Type 3 | Serial RS232 |
| Connector | DE-9 Female, Back Panel |
| Power Supply | |
| Line Voltage | 100 to 240 V |
| Line Frequency | 50 to 60 Hz |
| Maximum Power Consumption | 30W |
| Mains Supply Overvoltage | Category II (Cat II) |
| General | |
| Dimensions (W x H x D) | 253.2 mm x 110.6 mm x 293.6 mm (9.97" x 4.35" x 11.56") |
| Weight | 2.6 kg |
| Mounting Options | Benchtop |
| Operating Temperature Range | 0 to 40 °C |
| Storage Temperature Range | -40 to 70 °C |
| Relative Humidity | Max 80% up to 31 °C, Decreasing to 50% at 40 °C |
| Pollution Degree (Indoor Use Only) | 2 |
| Operation Altitude | <2000 m Above Sea Level |
Related Documents